
Referred to as residual chlorine, it is a water quality parameter for chlorine disinfection. The total amount of available chlorine remaining in water after the reaction of water with the added chlorine, in mg/L. When treating drinking water, chlorine or certain chlorine compounds (such as hypochlorite and chloramine compounds) are commonly used for disinfection. Sometimes chlorine is also used to oxidize the more polluted raw water to improve water quality. Residual chlorine is often determined to control the treatment process. The function of residual chlorine disinfectant in tap water treatment is to destroy bacteria, viruses and other microorganisms in the water, so as to ensure the health and safety of tap water. Chlorine disinfectants kill bacteria and viruses in water, preventing water contamination and the spread of infectious diseases. In tap water treatment, residual chlorine disinfectant is a commonly used disinfectant because of its good disinfection effect, easy to use and low cost. However, too high residual chlorine will bring odor to the water, and too low will make the water lose the ability to maintain sterilization, reducing the sanitary safety of water supply. This CL4.2 amperometric free chlorine sensor has a 4-20mA output and is suitable for chlorine measurement in drinking water and tap water. We offer different measuring ranges, which customers can choose according to their needs.
Features
2-electrode system, fast reaction speed
membrane-covered, amperometric, immersion measurement, data stability
4-20mA current output, RS485 module selectable
Internal automatic temperature compensation
Specification
Product Name | Chlorine Sensor |
Measure | Free inorganic chlorine |
Principle | membrane-covered, amperometric 2-electrode system |
PH Range | 6~8pH |
Pressure Range | 0~1 bar |
Temp Range | 0~45 ℃ |
TEMP Compensation | Automatic temperature compensation |
Response Time | T90 about 30s |
Flow Rate | About 30 l/h, low flow dependence |
Shaft Length | Standard 175 mm |
Connection | M12 male |
Output | 4-20mA |
Material | PVC-U, semipermeable membrane |
Application | Fresh water free of surfactants and with constant pH |
Calibration | At the controller, via analytical chlorine determination by DPD-1 method |
Interferences | ClO2, O3 |
If you are interested in the product, contact 2bvideo.com for more information for more information
The information of 2bvideo.com limited shown above is provided by the user or collected on the network. Video 2B does not guarantee the authenticity,accuracy and legitimacy of 2bvideo.com limited information. Video 2B does not involve legal relationships and disputes between users arising from transactions other than secured transactions on this website. Disputes shall be settled by you through negotiation. If you are the person in charge or relevant employee of this enterprise, if you find that the enterprise information is incorrect or want to manage thiscompany, please contact us jacklee1558@gmail.com, after you claim the enterprise, you can obtain management permission, publish supplyand demand information, bring consulting orders, and remove page advertisements.
Main Product:
Water Quality Monitor,
PH ORP Controller ,
PH ORP Analyzer,
PH Sensor,
Turbidity Sensor,
COD Sensor